Medical physics
Can you hear where a proton beam stops?
Not with your ears, but with sensitive acoustic sensors you can. A short proton pulse sets off a tiny pressure wave in water. From such signals, it is possible to reconstruct where the beam deposits its highest dose. A thin aluminium sheet placed in front of the tank shifted this spot – and the measurement showed it.
The peak just before the end
Protons are positively charged particles. As they travel through water or tissue, they lose energy. The energy absorbed per kilogram of material is called the dose. Shortly before the protons come to a stop, the dose reaches a maximum: the Bragg peak. Beyond it, the dose drops off steeply. So the maximum and the end of the range are not the same point.[1]
In proton therapy, the dose distribution is planned so that the target area is irradiated while the surrounding tissue is spared as much as possible. This requires knowing how far the protons travel. Calculation inaccuracies, the positioning of the patient and organ motion make this range uncertain, and they have to be accounted for during planning.[1] A measurement taken during irradiation could help verify the actual position.
From pulse to pressure wave
A very short proton pulse slightly warms the water where it deposits energy. The water expands, and a pressure wave is created. For the setup described, the researchers specify pulses shorter than 200 billionths of a second.[1]
In a similar way, lightning heats the air and thunder follows. From the delay, you can estimate how far away the storm is. In the water tank, the distances are tiny and the sound is in the ultrasound range. The reconstruction uses several recordings, known sensor positions and a model of how sound travels to determine where the wave originated. The method is called ionoacoustics.[1]
From a short pulse to a reconstructed position
Four schematic steps: a proton pulse warms a small zone, a pressure wave spreads out, sensors record it, and a computational model reconstructs its origin.
Pulse and heating
The short proton pulse deposits energy locally. The water warms very slightly and expands.
Pressure wave
The change in pressure spreads through the water from the excited region.
Recording
Ultrasound sensors pick up the wave. Their positions are known for the later reconstruction.
Locating the origin
The position of the origin is reconstructed from the recordings and a model of wave propagation.
One millimetre closer to the foil
A team led by Stephan Kellnberger sent 20 MeV protons through a thin plastic foil into a water tank. MeV is a unit of energy for individual particles. A curved array of 64 ultrasound sensors was moved step by step across the beam. From the signals, the team spatially reconstructed the position of the maximum and the shape of the Bragg peak.[1]
The maximum lay 4.3 ± 0.2 millimetres behind the foil. This value gives the position together with its reported uncertainty. It agreed with earlier measurements and computer simulations.[1]
Next, a 0.5-millimetre-thick aluminium sheet was placed in the beam before it reached the water. The protons lost energy in the sheet and travelled less far in the water. The maximum now lay 3.3 ± 0.2 millimetres behind the foil, about one millimetre closer.[1] The original figure shows this difference in panels c and e. It displays relative profiles, not an absolutely calibrated three-dimensional dose map.
Two positions in water

How to read the figure
- a shows the setup: ion beam is the incoming proton beam, US waves are the ultrasound waves, US array is the arrangement of ultrasound sensors, and water phantom is the water tank. xyz-stage refers to the device that moves the sensors in space.
- In c and e, the beam runs along z. Polyimide foil is the entrance foil at z = 0; air and water mark the regions of air and water. The x, y and z axes give distances in millimetres. b and d show projections across the beam direction.
- c: without aluminium, the maximum lies 4.3 ± 0.2 mm behind the foil. e: with the 0.5 mm aluminium sheet placed upstream, it lies at 3.3 ± 0.2 mm. The second value is given in the results section of the paper.
- dz marks the distance of the maximum from the foil. dr denotes the longitudinal width at half the maximum height: 0.28 ± 0.05 mm in c, 0.35 ± 0.05 mm in e. This width is neither a position uncertainty nor a deviation from a reference measurement.
- dose (a.u.) means dose in relative (arbitrary) units. The white profile line and the position of the peak show the shift even without relying on colour. The original axes and all five panels remain accessible in the enlarged view.
Figure 2: S. Kellnberger et al., Scientific Reports 6, 29305 (2016), DOI: 10.1038/srep29305. CC BY 4.0. Original unchanged; translated reading guide added.
Full original caption (English)
Three-dimensional tomographic scan of the Bragg peak. ( a ) Experimental setup for 3D ionoacoustic imaging. ( b ) Maximum intensity projection of the 3D reconstruction in the xy plane. The arrows indicate the position of the maximum in x and y axis. ( c ) Image of the reconstructed volume, showing the MIP in the yz -plane. The line profile depicts the position of the Bragg peak at a distance of 4.3 ± 0.2 mm from the polyimide foil and further reveals a longitudinal FWHM of the Bragg peak of 0.28 ± 0.05 mm. ( d ) Image of the Bragg peak after introducing an aluminum absorber in the beam path. The arrows indicate the position of the Bragg peak maximum. ( e ) Proton range determination using ionoacoustic tomography. The inserted aluminum absorber reduces the range of protons in water.
What the test at a clinical accelerator showed
At a facility used for therapy, a team led by Sebastian Lehrack studied protons with initial energies between 145 and 227 MeV. Here, too, the measurements were made in water. A sensitive underwater microphone, a hydrophone, recorded the sound. The acoustically determined ranges were compared with data from a calibrated measurement setup on site.[2]
At 220 MeV and about 10 gray at the Bragg peak, the two measurements agreed to within less than one millimetre, within the measurement errors.[2] The gray is the unit of dose, meaning the energy absorbed per kilogram. The 10 gray belong to this particular test condition; the millimetre figure describes the comparison in water.
What remains open inside the body
In the leg of a mouse that had been killed before the experiment, the Bragg peak could also be imaged. The researchers report a lower signal-to-noise ratio than in their previous experiments.[1] This is a single experiment performed after the animal's death, not proof that a measurement works during a treatment.
The authors of the clinical accelerator study see potential for submillimetre measurements during proton therapy. That is their outlook; the reported result comes from water.[2] Organ motion remains a separate problem for range planning.[1]
Walter Assmann, Sebastian Lehrack, Peter G. Thirolf and Katia Parodi were among those involved in both studies. The experiments at different facilities each contribute their own findings.[1][2] The key distinction remains: locating a peak in water has been demonstrated – reliable verification inside the living body requires further testing.
Sources
[1] S. Kellnberger, W. Assmann, S. Lehrack et al.: Ionoacoustic tomography of the proton Bragg peak in combination with ultrasound and optoacoustic imaging. Scientific Reports 6, 29305 (2016). DOI: 10.1038/srep29305.
Open original source [1][2] S. Lehrack, W. Assmann, D. Bertrand et al.: Submillimeter ionoacoustic range determination for protons in water at a clinical synchrocyclotron. Physics in Medicine & Biology 62(17), L20–L30 (2017). DOI: 10.1088/1361-6560/aa81f8.
Open original source [2]