Medical and health technology

Why can a heart rate watch be less accurate during walking than during cycling?

Share article

Why can a heart rate watch be less accurate during walking than during cycling?

In a 2017 laboratory comparison, heart rate watches deviated more from the ECG during walking than during cycling on a stationary cycle ergometer.[1] The reason was not examined as a separate cause. Movement at the wrist is one possible explanation from the research literature. A second study describes such sources of interference and also shows why this explanation has limits.[2]

What the watch measures at the wrist

Heart rate watches measure the pulse with light. Small light sources on their underside, usually green, shine into the skin, and a sensor detects the light that comes back. With each heartbeat, the amount of blood in the vessels beneath the skin changes, and with it how much light arrives at the sensor. From these rises and falls, the watch calculates the heart rate. The method is called photoplethysmography, or PPG for short.[2]

A small shift at the wrist

Two simplified cross-sections: an optical sensor rests steadily on the skin; next to it, the sensor is slightly shifted sideways and lifted. Light source and receiver sit above a schematic vascular zone.

Steady fit

Slightly shifted fit

  • Light source
  • Light receiver
  • Tissue and vascular zone

Circle: light source. Square: light receiver. Area: tissue. Branching line: simplified vascular zone.

Schematic created for this article, not measurement data: a change in contact can affect the optical signal. Sizes, spacing and displacement are not to scale; light paths and pulse waves are not shown.

How movement can disturb the light signal

When the arm moves, the light signal can contain interference. The authors of the second study name typical causes: the sensor shifting on the skin, the skin deforming, and changes in blood flow. The watch may then miss heartbeats or count extra ones.[2]

There is also an effect that earlier work calls “signal crossover”. During regularly repeated movements such as walking or jogging, motion signals and pulse signals can overlap. The analysis algorithm may then pick up the rhythm of the movement instead of the pulse.[2] This possible explanation comes from earlier research. The comparisons reported here do not isolate arm movement as the cause of the differences between activities.[1][2]

What the first laboratory test found

A team at Stanford University tested seven watches and wristbands available at the time on 60 volunteers. Each person wore up to four devices at once. An ECG, which records the electrical activity of the heart, ran alongside them and served as the reference. After a seated phase, participants walked and ran on a treadmill and then rode the ergometer, each at two levels. The last minute of each level was analyzed.[1]

The caption of Figure 2A defines the error as the absolute deviation divided by the ECG value: what counts is the size of the difference, not its direction. For the laboratory conditions examined there, the researchers report medians across devices of 1.8 percent for cycling and 5.5 percent for walking.[1] The median is the middle value when all deviations are sorted by size. A worked example, not a measurement from the study: at a pulse of 100 beats per minute, 5.5 percent would be about five to six beats.

The researchers describe cycling as the most stable of the activities tested. One finding, however, does not fit a simple explanation based on movement rhythm: deviations were smaller during running than during walking. Within the same activity, they increased with intensity.[1] Running is also a regular movement. Rhythm alone therefore does not explain the difference.

What the second laboratory test showed

A team at Duke University examined six optical devices in 53 people who completed the full protocol. Five devices were worn on the wrist, one on the upper arm. An ECG again served as the reference. Participants first sat quietly, then walked on a treadmill for five minutes and, after a break, typed on a keyboard for one minute. Cycling was not part of the protocol.[2]

The measure used here, the mean absolute error (MAE), is given in beats per minute and is likewise unsigned. During walking, it was higher than at rest for five of the six devices and stayed about the same for one. In Figure 3a, two devices are marked as statistically significantly different, not all of them.[2] During typing — a small, repetitive hand movement while seated — the mean absolute error was also above the resting value for several devices. This is shown in Figure 2B of the supplement.[2] This comparison alone does not yet establish which mechanism causes the difference.

Walking and rest compared

Excerpt from Figure 3, panel a: mean absolute heart rate error of six devices. Orange shows walking, turquoise shows rest. For five devices the walking value is higher; for Apple Watch it is about the same. Double asterisks mark Fitbit and Empatica.

How to read the figure

  • Orange: walking on the treadmill. Turquoise: seated rest. The colored dots and error bars belong to these two conditions.
  • The vertical axis shows the mean absolute error (MAE) between ECG and device in beats per minute. Lower means more accurate in this comparison, not better in general. The green horizontal line marks zero.
  • Error bars show 95% confidence intervals. In this panel, double asterisks indicate a significant difference for two devices: Fitbit and Empatica. Not every visible difference is statistically supported.
  • The blue group bars distinguish four consumer devices (Consumer) from two research devices (Research). Biovotion was worn on the upper arm, the other devices on the wrist. The graphic is not a ranking of today's watches.

Source: B. Bent, B. A. Goldstein, W. A. Kibbe and J. P. Dunn, npj Digital Medicine 3, 18 (2020), DOI: 10.1038/s41746-020-0226-6. CC BY 4.0. Modification: excerpt, panel a; original pixels and aspect ratio preserved.

Original caption for panel a (English)

a Mean absolute error in heart rate (bpm) across devices during rest (teal) and activity (orange). This shows the true difference in HR from the ECG but does not show the sign of the difference. The green horizontal line represents no error (no difference from the true measurement of HR from ECG). Error bars show the 95% confidence interval. ** indicates significant difference in error between baseline and activity with a Bonferroni multiple hypothesis corrected p value of 0.0042.

Walking and rest compared

The excerpt can be enlarged and panned horizontally and vertically within the image area. Press Escape to close; focus returns to the open button.

100%
Excerpt from Figure 3, panel a: mean absolute heart rate error of six devices. Orange shows walking, turquoise shows rest. For five devices the walking value is higher; for Apple Watch it is about the same. Double asterisks mark Fitbit and Empatica.

What the numbers do and do not tell us

The two studies use different measures: percent of the ECG value in the first, beats per minute in the second. They tested different devices in different protocols, so the values cannot be directly combined. Both concerned devices and software of that time. They do not yield a ranking of today's watches.[1][2]

In both experiments, the deviation depended on what participants were doing at the time. How much of this is due to movement at the wrist remains open. When looking at a test result for a heart rate watch, it is therefore worth checking which activity it was measured during.[1][2]

Sources

[1] A. Shcherbina, C. M. Mattsson, D. Waggott et al.: Accuracy in Wrist-Worn, Sensor-Based Measurements of Heart Rate and Energy Expenditure in a Diverse Cohort. Journal of Personalized Medicine 7(2), 2017, 3.

Open original source [1]

[2] B. Bent, B. A. Goldstein, W. A. Kibbe, J. P. Dunn: Investigating sources of inaccuracy in wearable optical heart rate sensors. npj Digital Medicine 3, 2020, 18.

Open original source [2]